United States Government Work, uncopyrighted, public-domain, DRAFT COPY ONLY. This document does not necessarily represent the views or policies of any United States Government agency. This document is provided "as is", without warranty of any kind, express or implied, including but not limited to the warranties of merchantability, fitness for a particular purpose and non-infringement. In no event shall the authors be liable for any claim, damages or other liability, whether in an action of contract, tort or otherwise, arising from, out of, or in connection with the document or the use or other dealings made with the document.

See also: Tissue-Tek® Processor: 11, 13, 14.



PRINCIPLE OF THE TEST.


Obtaining processed tissue from the TISSUE TEK VIP®5 Vacuum Infiltration Processor.

SPECIMEN REQUIRED.


All human tissue excised at surgery, outpatient clinics, and postmortems, fresh or in fixative, along with a filled-out Tissue Examination Form (U. S. Standard Form 515, USSF515).



REAGENTS, INSTRUMENTATION.


TISSUE TEK VIP®5 Vacuum Infiltration Processor.



STEP-BY-STEP DESCRIPTION.


1. End of process push drain.

2. Press drain button to drain paraffin.

3. Wait about five minutes, or until paraffin drains out of the chamber or reservoir (pot on top of machine).

4. Lift top of pot, and take square basket out of reservoir pot; place basket with tissue intact on top of machine. Cassettes will harden but can be remelted down the following day.

5. Wait a few minutes (5 to 10 minutes), and push button to purge; the machine is ready for next daily run.

6. These instructions are also posted on the front of the machine.

7. Do not turn the power off the processor.



REFERENCE.


1. Sakura Tissue-Tek® VIP® 5 Vacuum Infiltration Processor. Operating Manual.
© 2001. Sakura Finitek, U.S.A., Inc.
Technical Support: 1-800-543-8496.
Sakura Finitek, U.S.A., Inc. © 2001.
1750 W. 214th Street
Torrance, CA 90501